7
General Maintenance
To Open the Analyzer Chamber
176
5977B Series MSD Operating Manual
To Open the Analyzer Chamber
The analyzer chamber should only be opened to clean or replace the ion source,
change the detector’s EM, or to change a filament.
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Wrist strap, antistatic
•
Small (9300-0969)
•
Medium (9300-1257)
•
Large (9300-0970)
Procedure
1
2
Open the left side panel. (See
3
Loosen the analyzer side plate thumbscrews if they are fastened. (See
CAUTION
Electrostatic discharges to analyzer components are conducted to the quad
driver board, where they can damage sensitive components. Wear a grounded
antistatic wrist strap and take other antistatic precautions (See
on page 171.) before you open the analyzer chamber.
WARNING
The analyzer, GC/MSD interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
Содержание 5977B CI MSD
Страница 1: ...Agilent 5977B Series MSD Operating Manual ...
Страница 12: ...12 5977B Series MSD Operating Manual ...
Страница 34: ...1 Introduction To Replace the Primary Fuses 34 5977B Series MSD Operating Manual ...
Страница 54: ...2 Installing 8890 GC Columns The GC MSD Interface for an 8890 Series GC 54 5977B Series MSD Operating Manual ...
Страница 100: ...4 Installing 7890 GC Columns The GC MSD Interface for a 7890B Series GC 100 5977B Series MSD Operating Manual ...
Страница 142: ...5 Operating in EI Mode To Pumpdown the MSD 142 5977B Series MSD Operating Manual ...
Страница 252: ...8 CI Maintenance To Install the CI Source 252 5977B Series MSD Operating Manual ...
Страница 253: ......
Страница 254: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G7077 90034 G7077 90034 ...