93
Chapter 4 Features and Functions
Scanning
4
Power Failure
• When shipped from the factory, the instrument is configured to
automatically recall the power-down state when power is restored. In
this configuration, the instrument will automatically recall the
instrument state at power-down and resume a scan in progress. If you
do not want the power-down state to be recalled when power is
restored, send the
MEMory:STATe:RECall:AUTO OFF
command
(also see the
Utility
menu); a Factory Reset (
*RST
command) is then
issued when power is restored.
• If the instrument is in the middle of a scan sweep when power fails,
all readings from that partially completed sweep will be discarded (a
sweep
is one pass through the scan list). For example, assume that
your scan list includes four multiplexer channels and you want to
sweep through the scan list three times (see diagram). A power
failure occurs after the second reading in the third scan sweep. The
instrument will discard the last two of the 10 readings and will
resume scanning at the beginning of the third scan sweep.
• If you remove a module or move a module to a different slot while
power is off, the scan
will not
resume when power is restored. No
error is generated.
• If you replace a module with a module of the same type while power is
off, the instrument will continue scanning when power is restored. No
error is generated.
Power Fails
34970A Refresh UG.book Page 93 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...