227
Chapter 5 Error Messages
Instrument Errors
4
5
281
Not able to perform on more than one channe
l
You can perform this operation on only one channel at a time. Check the
channel list that you sent with this command to see if it contains more
than one channel. This error is generated by the
ROUTe:MON
and
DATA:LAST?
commands.
291
Not able to recall state: it is empty
You can only recall a state from a location that contains a previously
stored state. The state location that you attempted to recall is empty.
The storage locations are numbered 0 through 5.
292
Not able to recall state: DMM enable changed
The enable/disable state of the internal DMM has been changed since
the instrument state was stored. Use the
INSTrument:DMM?
command
to determine the state of the internal DMM.
For more information, see
“Internal DMM Disable” on page 167.
301
Module currently committed to scan
When you add a multiplexer channel to a scan list, that entire module is
dedicated to the scan. You cannot perform low-level close or open
operations on any channels on that module (even those channels that are
not configured). To stop a scan in progress, send the
ABORt
command or
a bus Device Clear.
303
Module not able to perform requested operation
A command was received which is not valid for the specified module.
This error is most commonly generated when you send a command
intended for the multifunction module to a switching module.
305
Not able to perform requested operation
The requested operation is not valid for the specified channel. You may
have tried to a configure a channel for current measurements (valid only
on channels 21 and 22 on the 34901A module). Or you may have tried to
configure scaling on a module that does not connect to the internal
DMM.
34970A Refresh UG.book Page 227 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...