142
Chapter 4 Features and Functions
Alarm Limits
• In addition to being stored in reading memory, alarms are also
recorded in their own SCPI status system. You can configure the
instrument to use the status system to generate a Service Request
(SRQ) when alarms are generated.
See the Agilent 34970A/34972A
Programmer’s Reference Help for more information.
• The default values for the upper and lower alarm limits are “0”. The
lower limit must
always
be less than or equal to the upper limit, even
if you are using only one of the limits.
• For details on configuring alarms on the multifunction module, see
“Using Alarms With the Multifunction Module” on page 148.
• A Factory Reset (
*RST
command) clears all alarm limits and turns
off all alarms. An Instrument Preset (
SYSTem:PRESet
command) or
Card Reset (
SYSTem:CPON
command)
does not
clear the alarm limits
and
does not
turn off alarms.
•
Front-Panel Operation
: To select the alarm for use on the active
channel, choose from the following items.
NO ALARM , USE ALARM 1 , ... USE ALARM 4
Then, choose from the following alarm conditions.
HI+LO ALARMS , HI ALARM ONLY , LO ALARM ONLY
Then, set the desired limit values and exit the menu. Note that the
instrument does not start evaluating the alarm conditions until you
exit the
Alarm
menu.
•
Remote Interface Operation
: To assign the alarm number to report
any alarm conditions on the specified channels, use the following
command (if not assigned, all alarms on all channels are reported on
Alarm 1 by default).
OUTPUT:ALARM2:SOURCE (@103,212)
34970A Refresh UG.book Page 142 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...