271
Chapter 7 Tutorial
Measurement Fundamentals
4
7
To make a more accurate measurement, you should extend the copper
test leads of the internal DMM closer to the measurement and hold the
connections to the thermocouple at the same temperature.
This circuit will give accurate temperature measurements. However, it is
not very convenient to make two thermocouple connections and keep all
connections at a known temperature. The
Law of Intermediate Metals
eliminates the need for the extra connection. This empirical law states
that a third metal (iron (Fe) in this example) inserted between two
dissimilar metals will have
no effect
upon the output voltage, provided
the junctions formed are at the same temperature. Removing the
reference thermocouple makes the connections much easier.
This circuit is the best solution for accurate thermocouple connections.
Measurement
Thermocouple
Reference Thermocouple
Ice Bath
Internal DMM
Internal DMM
Measurement
Thermocouple
Ice Bath
(External Reference Junction)
34970A Refresh UG.book Page 271 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...