Chapter 5 Error Messages
Instrument Errors
4
5
203
Memory lost: stored readings
This error is reported at power-on to indicate that readings stored in
memory from a previous scan have been lost. This error is most likely
caused by a dead battery (memory is battery-backed).
Refer to the
34970A/34972A Service Guide to replace the internal battery.
204
Memory lost: time and date
This error is reported at power-on to indicate that the time and date
settings have been lost (they are reset to
JAN 1, 1996 00:00:00
). This
error is most likely caused by a dead battery (memory is battery-backed).
Refer to the 34970A/34972A Service Guide to replace the internal battery.
221
Settings conflict: calculate limit state forced off
If you plan to use scaling on a channel which will also use alarms,
be sure
to configure the scaling values first
. This error is generated if you
attempt to assign the alarm limits first and the instrument will turn off
alarms and clear the limit values.
222
Settings conflict: module type does not match stored state
Before recalling a stored state, the instrument verifies that the same
module types are installed in each slot. The instrument has detected a
different module type in one or more slots.
223
Settings conflict: trig source changed to IMM
This error is generated if you attempt to set the channel advance source
(
ROUTe:CHAN:ADVance:SOURce
command) to the same source used for
the scan trigger (
TRIGger:SOURce
command). The command has been
accepted and executed but the scan trigger source is reset to
“
IMMediate
”.
224
Settings conflict: chan adv source changed to IMM
This error is generated if you attempt to set the scan trigger source
(
TRIGger:SOUR
ce command) to the same source used for the channel
advance source (
ROUTe:CHAN:ADVance:SOURc
e command). The
command has been accepted and executed but the channel advance
source is reset to “
IMMediate
”.
34970A Refresh UG.book Page 225 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...