222
Chapter 5 Error Messages
Execution Errors
-214
Trigger deadlock
A trigger deadlock occurs when the trigger source is
“BUS”
and a
READ?
command is received.
-221
Settings conflict
An invalid configuration was requested. This error is most commonly
generated when setting alarm limits. Note that the lower limit must
always be less than or equal to the upper limit, even if you are using only
one of the limits. This error is also generated if you send the
MEASure?
or
CONFigure
command with autorange enabled with a fixed resolution.
-222
Data out of range
A numeric parameter value is outside the valid range for this command.
Example
:
TRIG:COUNT -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 12 characters. This error can be generated
by the
CAL:STRing
and
DISPlay:TEXT
commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example
:
TRIG:SOURCE ALARM
(
ALARM
is not a valid choice)
-230
Data stale
A
FETCh?
or
DATA:REMove?
command was received but internal
reading memory was empty. The readings retrieved may be invalid.
-310
System error
A firmware defect has been found.
This is not a fatal error but you should
contact your nearest Agilent Service Center if this error is reported.
34970A Refresh UG.book Page 222 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...