263
Chapter 7 Tutorial
System Cabling and Connections
4
7
Noise Caused by Ground Loops
When measuring voltages in circuits
where the internal DMM and the device-under-test are both referenced
to a common earth ground, a
ground loop
is formed. As shown below, any
voltage difference between the two ground reference points (V
ground
)
causes a current to flow through the LO measurement lead. This causes
an error voltage (V
L
) which is added to the measured voltage.
Where:
R
L
= Lead resistance
R
i
= DMM isolation resistance
C
i
= DMM isolation capacitance
V
ground
= Ground noise voltage
I = Current flow caused by V
ground
=
Z
Z
Ci
=
10 M
at 50 or 60 Hz
V
L
=
To minimize ground loop errors:
• If V
ground
is a DC
voltage
, keep R
L
small compared to R
i
.
• If V
ground
is an AC
voltage
, keep R
L
small compared to Z, and set the
DMM’s integration time to 1 PLC or greater (
see page 120 for a
discussion of integration time
).
DMM
HI
LO
R
L
R
L
V
ground
I
V
test
C
i
250 pF
R
i
>10 G
V
ground
R
L
Z
+
---------------------
1
2
fC
-------------
I R
L
34970A Refresh UG.book Page 263 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...