Chapter 4 Features and Functions
Alarm Limits
4
• To set the upper and lower alarm limits on the specified channels, use
the following commands.
CALC:LIMIT:UPPER 5.25,(@103,212)
CALC:LIMIT:LOWER 0.025,(@103,212)
• To enable the upper and lower alarm limits on the specified channels,
use the following commands.
CALC:LIMIT:UPPER:STATE ON,(@103,212)
CALC:LIMIT:LOWER:STATE ON,(@103,212)
Viewing Stored Alarm Data
If an alarm occurs on a channel as it is being scanned, then that
channel’s alarm status is stored in
reading memory
as the readings are
taken. As alarm events are generated, they are also logged in an
alarm
queue
, which is separate from reading memory. This is the
only
place
where non-scanned alarms get logged (alarms during a monitor, alarms
generated by the multifunction module, etc.).
• You can store up to 50,000 readings in memory during a scan. You
can read the contents of reading memory at any time, even during a
scan. Reading memory
is not
cleared when you read it.
• Each time you start a new scan, the instrument clears all readings
(including alarm data) stored in reading memory from the previous
scan. Therefore, the contents of memory are always from the most
recent scan.
• Up to 20 alarms can be logged in the alarm queue. If more than 20
alarms are generated, they will be lost (only the first 20 are saved).
• The alarm queue is cleared by the
*CLS
(clear status) command,
when power is cycled, and by reading all of the entries. A Factory
Reset (
*RST
command) or Instrument Preset (
SYSTem:PRESet
command)
does not
clear the alarm queue.
•
Front-Panel Operation
: From the front panel, you can view the first
20 alarms in the queue. After turning the knob to the desired
channel, press
and
to view either the alarm reading or the
time that the alarm occurred. Notice that the annunciators indicate
which alarm is being viewed.
ALARMS
Note
: The alarm queue is cleared when you read the alarms.
34970A Refresh UG.book Page 143 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...