236
Chapter 5 Error Messages
Calibration Errors
Calibration Errors
The following errors indicate failures that may occur during a
calibration. Refer to the
34970A/34972A Service Guide
for more
information.
701
Cal: security disabled by jumper
The calibration security feature has been disabled with a jumper inside
the instrument. When applicable, this error will occur at power-on to
alert you that the instrument is unsecured.
702
Cal: secured
The instrument is secured against calibration.
703
Cal: invalid secure code
You have entered an invalid calibration security code. You must use the
same security code to unsecure the instrument that was used to secure
it, and vice versa. The security code may contain up to 12 alphanumeric
characters. The first character
must
be a letter, but the remaining
characters can be letters, numbers, or an underscore ( _ ). You do not
have to use all 12 characters but the first character must always be a
letter.
The security code is set to “HP034970” or “AT034972” when the
instrument is shipped from the factory.
704
Cal: secure code too long
The security code may contain up to 12 alphanumeric characters. A
security code was received which contained more than 12 characters.
705
Cal: aborted
A calibration in progress is aborted when you turn off the instrument or
send a bus Device Clear.
706
Cal: value out of range
The specified calibration value (
CALibration:VALue
) is not valid for
the present measurement function and range.
707
Cal: signal measurement out of range
The specified calibration value (
CALibration:VALue
) does not match
the signal applied to the instrument.
708
Cal: signal frequency out of range
The input signal frequency for an ac calibration does not match the
required input frequency for this calibration.
709
Cal: no cal for this function or range
You cannot perform calibrations for most ac current ranges, the 100 M
resistance range, and period.
34970A Refresh UG.book Page 236 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...