262
Chapter 7 Tutorial
System Cabling and Connections
Thermal EMF Errors
Thermoelectric voltages are the most common
source of error in low-level DC voltage measurements. Thermoelectric
voltages are generated when you make circuit connections using
dissimilar metals at different temperatures. Each metal-to-metal
junction forms a
thermocouple
, which generates a voltage proportional to
the junction temperature difference. You should take the necessary
precautions to minimize thermocouple voltages and temperature
variations in low-level voltage measurements. The best connections are
formed using copper-to-copper crimped connections. The table below
shows common thermoelectric voltages for connections between
dissimilar metals.
Noise Caused by Magnetic Fields
If you are making measurements
near magnetic fields, you should take precautions to avoid inducing
voltages in the measurement connections. Voltage can be induced by
either movement of the input connection wiring in a fixed magnetic field
or by a varying magnetic field. An unshielded, poorly dressed input wire
moving in the earth’s magnetic field can generate several millivolts. The
varying magnetic field around the AC power line can also induce
voltages up to several hundred millivolts. You should be especially
careful when working near conductors carrying large currents.
Where possible, you should route cabling away from magnetic fields.
Magnetic fields are commonly present around electric motors,
generators, televisions, and computer monitors. Also make sure that
your input wiring has proper strain relief and is tied down securely when
operating near magnetic fields. Use twisted-pair connections to the
instrument to reduce the noise pickup loop area, or dress the wires as
close together as possible.
Copper-to-
Approx.
V/°C
Copper
Gold
Silver
Brass
Beryllium Copper
Aluminum
Kovar or Alloy 42
Silicon
Copper-Oxide
Cadmium-Tin Solder
Tim-Lead Solder
<0.3
0.5
0.5
3
5
5
40
500
1000
0.2
5
34970A Refresh UG.book Page 262 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...