Chapter 4 Features and Functions
System-Related Operations
System-Related Operations
This section gives information on system-related topics such as storing
instrument states, reading errors, running a self-test, displaying
messages on the front panel, setting the system clock, disabling the
internal DMM, reading the firmware revisions, and reading the relay
cycle count.
State Storage
The instrument has six storage locations in non-volatile memory to store
instrument states. The locations are numbered 0 through 5. The
instrument uses location “0” to automatically hold the state of the
instrument at power down. You can also assign a name to each of the
locations (1 through 5) for use from the front panel.
• You can store the instrument state in any of the six locations.
However, you can only recall a state from a location that contains a
previously stored state. You can use location “0” to store a sixth
instrument state. However, keep in mind that location “0” is
automatically overwritten when power is cycled.
• The instrument stores the state of all modules including all channel
configurations, scanning setups, alarm values, and scaling values.
• When shipped from the factory, storage locations “1” through “5” are
empty (location “0” has the power-on state).
• When shipped from the factory, the instrument is configured to
automatically recall the power-down state (state “0”) when power is
restored. You can change the factory configuration such that a
Factory Reset (
*RST
command) is issued when power is restored.
• Before recalling a stored state, the instrument verifies that the same
module types are installed in each slot. If a different module type is
installed, the instrument will perform the equivalent of a Card Reset
(
SYSTem:CPON
command) on that slot.
34970A Refresh UG.book Page 160 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...