130
Chapter 4 Features and Functions
Voltage Measurement Configuration
Voltage Measurement Configuration
To connect voltage sources to the module’s screw terminals, see page 28.
This section contains information to help you configure the instrument
for making voltage measurements. The instrument can measure DC and
true RMS ac-coupled voltages on the measurement ranges shown below.
DC Input Resistance
Normally, the instrument’s input resistance is fixed at 10 M
for all DC
voltage ranges to minimize noise pickup. To reduce the effects of
measurement loading errors, you can set the input resistance to greater
than 10 G
for the 100 mVDC, 1 VDC, and 10 VDC ranges.
Applies to DC voltage measurements only.
• The instrument selects 10 M
(fixed input resistance on all DC
voltage ranges) when the measurement function is changed or after a
Factory Reset (
*RST
command). An Instrument Preset
(
SYSTem:PRESet
command) or Card Reset (
SYSTem:CPON
command)
does not
change the input resistance setting.
•
Front-Panel Operation
: First, select the DC voltage function on the
active channel. Then, go to the
Advanced
menu and select 10 M
(fixed resistance for all DC voltage ranges) or >10 G
.
The default is
10 M
INPUT R >10 G
100 mV
1V
10V
100V
300V
Autorange
Input Resistance
Setting
Input Resistance
100 mV, 1V, 10 V ranges
Input Resistance
100 V, 300 V ranges
Input R Auto OFF
Input R Auto ON
10 M
> 10 G
10 M
10 M
34970A Refresh UG.book Page 130 Wednesday, February 17, 2010 12:34 PM
Содержание 34970A
Страница 16: ...16 34970A Refresh UG book Page 16 Wednesday February 17 2010 12 34 PM ...
Страница 20: ...34970A Refresh UG book Page 20 Wednesday February 17 2010 12 34 PM ...
Страница 21: ...1 1 Quick Start 34970A Refresh UG book Page 21 Wednesday February 17 2010 12 34 PM ...
Страница 39: ...2 2 Front Panel Overview 34970A Refresh UG book Page 39 Wednesday February 17 2010 12 34 PM ...
Страница 59: ...3 3 System Overview 34970A Refresh UG book Page 59 Wednesday February 17 2010 12 34 PM ...
Страница 87: ...4 4 Features and Functions 34970A Refresh UG book Page 87 Wednesday February 17 2010 12 34 PM ...
Страница 217: ...5 5 Error Messages 34970A Refresh UG book Page 217 Wednesday February 17 2010 12 34 PM ...
Страница 240: ...240 Chapter 5 Error Messages Plug In Module Errors 34970A Refresh UG book Page 240 Wednesday February 17 2010 12 34 PM ...
Страница 241: ...6 6 Application Programs 34970A Refresh UG book Page 241 Wednesday February 17 2010 12 34 PM ...
Страница 254: ...254 34970A Refresh UG book Page 254 Wednesday February 17 2010 12 34 PM ...
Страница 255: ...7 7 Tutorial 34970A Refresh UG book Page 255 Wednesday February 17 2010 12 34 PM ...
Страница 324: ...324 34970A Refresh UG book Page 324 Wednesday February 17 2010 12 34 PM ...