Introduction to Programming
4
Agilent 34945A Microwave Switch/Attenuator Driver Module User’s Guide
87
Table 4-1. 34945A Factory Settings.
Parameter
Command
Setting
34945EXT Settings
Channel Drive Source
ROUTe:RMODule:DRIVe:SOURce:IMMediate
OFF
Boot/Reset Channel Drive
ROUTe:RMODule:DRIVe:SOURce:BOOT
OFF
Simultaneous Operations
ROUTe:RMODule:DRIVe:LIMit
1
Bank Settings
Channel Drive Type
ROUTe:RMODule:BANK:DRIVe:MODE
OCOLlector
LED Drives
ROUTe:RMODule:BANK:LED:DRIVe:ENABle
ON
LED Drive Level
ROUTe:RMODule:BANK:LED:DRIVe:LEVel
5 mA
Channel Settings
Operation Mode
ROUTe:CHANnel:DRIVe:PAIRed:MODE
OFF
Actuation Type
ROUTe:CHANnel:DRIVe:PULSe:MODE
ON
Pulse Width
ROUTe:CHANnel:DRIVe:PULSe:WIDTh
15 ms
Recovery Time
ROUTe:CHANnel:DRIVe:TIME:RECovery
0.0 seconds
Settle Time
ROUTe:CHANnel:DRIVe:TIME:SETTle
0.0 seconds
Channel State
(after boot or reset)
ROUTe:CHANnel:DRIVe:OPEN:DEFault
OPEN
Verification
ROUTe:CHANnel:VERify:ENABle
OFF
Verification Polarity
ROUTe:CHANnel:VERify:POLarity
NORMal
Содержание 34945A
Страница 8: ...viii Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide ...
Страница 22: ...14 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 1 About the Instrument ...
Страница 96: ...88 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 4 Introduction to Programming ...
Страница 120: ...112 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 5 Additional Operating Information ...