54
Agilent 34945A Microwave Switch/Attenuator Driver Module User’s Guide
3
Software Installation and Configuration
Figure 3-11. Specifying the Instrument Address.
4. The update process takes several minutes to complete. The 34980A front
panel will display “
UPDATING FIRMWARE
” during this time. DO NOT cycle power
during the process.
Following the update the 34980A will reboot. Once the reboot sequence com-
pletes, cycle power. To view the revision number of the installed update from
the 34980A front panel, press ‘UTILITY’ and rotate the front panel knob until
‘SELECT MENU FIRMWARE’ is displayed. Press ‘ENTER’ once and with
“REVISIONS” highlighted, press ‘ENTER’ a second time. Rotating the knob will
display the firmware revisions of the mainframe, front panel, DMM (if
present), and installed modules (e.g. 34945A) by slot number.
To verify that the 34945EXT has the same firmware revision as the 34945A,
the following commands can be sent from the Interactive IO utility within Agi-
lent Configuration Expert:
SYSTem:CTYPe? <slot>
!returns firmware revision of the 34945A
SYSTem:CTYPe:RMODule? (@<rem_ch>) [,<dist_board>] !returns firmware
revision of the
34945EXT
For example, if the 34945A is installed in 34980A slot 2,
S
YST:CYPE? 2
returns
the serial number and firmware revision of the 34945A. Sending
SYST:CTYPE:RMOD? (@2100)
returns the serial number and firmware revision
of the 34945EXT. The revision numbers must be the same.
Содержание 34945A
Страница 8: ...viii Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide ...
Страница 22: ...14 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 1 About the Instrument ...
Страница 96: ...88 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 4 Introduction to Programming ...
Страница 120: ...112 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 5 Additional Operating Information ...
Страница 124: ...116 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide Index ...