106
Agilent 34945A Microwave Switch/Attenuator Driver Module User’s Guide
5
Additional Operating Information
Figure 5-6 illustrates the pulse actuation for two channels (switches) and the
relationship of the drive parameters to the power supply requirements.
Figure 5-6. Pulse Actuation Mode Diagram.
As shown in the figure , the actuation pulse is applied to channel 1 and held
for the T(Pulse) time set by the bank
preset
which is 15.0 ms, or as set with the
command:
ROUTe:CHANnel:DRIVe:PULSe:WIDTh
The actuation pulse is applied to channel 2 only after a power supply recovery
period has elapsed T(Recovery). The power supply recovery time can be set for
each channel, and is set to 0.0 ms during a bank
preset
(Chapter 4). The
recovery time can be changed using the command:
ROUTe:CHANnel:DRIVe:TIMe:RECovery
When verifying channel closure, you can also specify a settling time (T(Settle)).
This parameter ensures the switch has had time to change state before the
position indicator (of the switch) is evaluated.
The settling time can be set for each channel and is set to 0.0 ms during a bank
preset
(Chapter 4). The settling time can be changed using the command
ROUTe:CHANnel:DRIVe:TIMe:SETTle
The 34945A will always pulse a channel in response to a
ROUTe:OPEN
or
ROUTe:CLOSe
command. For example, sending
ROUTe:CLOSe
to a channel three
times in a row will result in three actuation pulses.
T(Setttle)
T(Recovery)
Drive Ch 1
Drive Ch2
Start Drive
Channel 1
Start Drive
Channel 2
Channel 1 Position
Indicators Evaluated
T(Pulse)
Содержание 34945A
Страница 8: ...viii Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide ...
Страница 22: ...14 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 1 About the Instrument ...
Страница 96: ...88 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 4 Introduction to Programming ...
Страница 120: ...112 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 5 Additional Operating Information ...
Страница 124: ...116 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide Index ...