Agilent 34945A Microwave Switch/Attenuator Driver Module User’s Guide
5
Additional Operating Information
Figure 5-3. Single-Coil Switch Connection Diagram (separate position indicators).
Using Dual-Coil Switches and Attenuators
Many microwave
switches and attenuators have a paired drive input.
Typically, one drive is electrically connected to the lower channel number in a
bank and one connected to a corresponding upper channel number. For
example, on bank 2 a dual drive switch should have its ‘State A’ coil connected
to channel 21 and its ‘State B’ coil connected to channel 31.
The 34945A drives dual coil devices in pulsed mode only
(
ROUTe:CHANnel:DRIVe:PULSe:MODE ON
). Pairing two channels
(
ROUTe:CHANnel:DRIVe:PAIR:MODE ON
) automatically configures the
channels to pulsed mode. You must turn the paired mode ‘off’ before
continuous drive mode can be enabled).
Based on the distribution board installed,
presetting
the bank will either
enable paired mode (ON) or disable paired mode (OFF) according to the type
of switches supported by the board.
34945EXT
Y1155A
Distribution
Board
Switch
6
5
4
3
2
1
1
3
4
5
6
2
Open
All
IND 1
+VI
+VR
DRV 1
Logic Gate
Sense
Pull Down
Resistor
Open Collector
Output Driver
To DRV 2 through 6
To DRV 7
To IND 2 through 6
Содержание 34945A
Страница 8: ...viii Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide ...
Страница 22: ...14 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 1 About the Instrument ...
Страница 96: ...88 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 4 Introduction to Programming ...
Страница 120: ...112 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide 5 Additional Operating Information ...
Страница 124: ...116 Agilent 34945A Microwave Switch Attenuator Driver Module User s Guide Index ...