![Advanced Instruments 4C3 Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/advanced-instruments/4c3/4c3_user-manual_2856032070.webp)
“[START] Ready?”
, using the techniques recommended in
Chapter 2, pipet five samples each of -408 m°C (-422 m°H) and
-600 m°C (-621 m°H) Advanced Cryoscope Standard, place them,
in that order, in a cassette and place the cassette in the left (input)
deck (see figure 5). When ready, press
“
START
”
to continue.
2.
Using an automatic calibration cassette:
By means of the techniques recommended in Chapter 2, pipet five
samples each of -408 m°C (-422 m°H) and -600 m°C (-621 m°H)
standard, place them, in that order, in an Advanced automatic cali-
bration cassette and place the cassette in the left deck.
If testing is not currently in process, press
“
START
”
. When the 4C3
identifies the Automatic Calibration Cassette, it will verify that
position 1 appears to contain the low standard, test the 10 samples
in the cassette, automatically recalibrate, then resume regular testing
on the samples in any following cassettes.
NOTES
a. If the supervisor/operator keyswitch is not in the super-
visor (vertical) position when the CALIB button is
pressed or an automatic calibration cassette is introduced,
“Supervisor Key Needed”
will be displayed and the
instrument will await operator response. Switch to super-
visor to continue or press STOP to cancel re-calibration.
b. If the sample probe is replaced, the sample probe bin
#
must be determined and set before the 4C3 can
be calibrated (see Chapter 5). After the sample probe
bin
number is set, the CALIB calibration procedure
(choice 1) must be used; choice 2 will generally produce
a
“Calibrator out of Range”
error message.
c. The 4C3 can detect certain gross calibration errors such
as reversal of the standards, but it remains the responsi-
bility of the operator to be sure to introduce accurate stan-
dards of the specified freezing point values.
d. The ambient air may contaminate or evaporate samples in
the cassette if they are allowed to remain for any appre-
48
The Advanced
®
Model 4C3 User’s Guide
Содержание 4C3
Страница 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Страница 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Страница 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Страница 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Страница 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Страница 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Страница 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Страница 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Страница 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Страница 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...