Troubleshooting Table
“System Error:
Requests”
“System Error:
Primitive”
“System
Error: Interrupt”
“System Error:
Communications”
“System Error:
Unknown”
“System Error:
Trap”
“System Error:
NMI”
“System Error:
ID”
Switch the instrument
off, then on.
Retry the test.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
“Stir/Freeze
Driver Failure”
Switch the instrument
off, then on.
Retry the test.
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Switch the instrument
off, then on.
Check the sample probe
by running the A/D
Test (see Chapter 5,
Section G.1).
If the error message
does not persist and
other messages are not
displayed, ignore this
message.
Replace the sample
probe (see Chapter 5,
Section H.3).
“Event
Record Lost”
Symptom
Check
Suggested Action
Содержание 4C3
Страница 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Страница 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Страница 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Страница 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Страница 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Страница 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Страница 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Страница 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Страница 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Страница 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...