The Advanced
®
Model 4C3 User’s Guide
“Head
Rotation Error”
“Head Not
Up” or “Head
Not Down”
Switch the instrument off,
then on.
Check the operating head
alignment by running the
Head Up/Down Test (see
Chapter 5, Section G).
Check the probe and stir/
freeze wire alignment
(see Chapter 5, Section
H).
Visually check for a
sample tube stuck on the
mandrel or other similar
mechanical cause.
a. Clear and discard the
stuck sample tube or
other visible cause of
the malfunction.
b. Check that the probe
and stir/freeze wire are
still in proper position
(see Chapter 5, Section
H).
c. Press STOP to reset the
instrument.
d. Remove the interrupted
cassette. To finish test-
ing it, remove already
tested samples and
place the cassette all
the way to the left in
the input deck.
e. Press START to con-
tinue testing. If the
problem recurs, see
Chapter 5, Section C
for further assistance.
Re-align the probe
and stir/freeze wire as
necessary (see Chapter
5, Section H).
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Symptom
Check
Suggested Action
Содержание 4C3
Страница 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Страница 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Страница 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Страница 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Страница 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Страница 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Страница 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Страница 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Страница 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Страница 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...