Troubleshooting Table
“Duplicate
Cassette #”
“Truncated
Barcode”
Results not
repeatable
(too scattered).
“No Barcodes
for This
Cassette”
Cassette. Cassettes cannot
be re-introduced during
Batch or Barcode runs.
Use a cassette not previ-
ously used in this run or
start another batch.
Cassette. If an un-scanned
cassette is introduced
during a scanned batch
run, the cassette will be
tested and the results
recorded without batch
identification.
Try to avoid introduc-
ing unscanned cassettes
during batch test runs.
Try to avoid the use of
barcodes longer than
16 characters.
Truncated barcodes lon-
ger than 16 characters
may appear as duplicate
entries.
1. Operator procedures
and technique.
2. Check the sample
probe by running the
A/D test (see Chapter
5, Section G).
“Fluid Level =
Error”
A wiring or circuit board
error is indicated. If the
error message persists, see
Chapter 5, Section C for
for further assistance.
Symptom
Check
Suggested Action
See Chapter 2,
Sections D and G.
Replace the sample
probe (see Chapter 4,
Section G).
If the error message
persists, see Chapter 5,
Section C for further
assistance.
Содержание 4C3
Страница 1: ...The Advanced Cryoscope Model 4C3 User s Guide Part No 4C35 Rev17 042318 ...
Страница 6: ...vi The Advanced Model 4C3 User s Guide Notes ...
Страница 10: ...x The Advanced Model 4C3 User s Guide Notes ...
Страница 14: ...Notes xiv The Advanced Model 4C3 User s Guide ...
Страница 68: ...46 The Advanced Model 4C3 User s Guide Notes ...
Страница 116: ...94 The Advanced Model 4C3 User s Guide Notes ...
Страница 122: ...100 The Advanced Model 4C3 User s Guide Notes ...
Страница 126: ...104 The Advanced Model 4C3 User s Guide Notes ...
Страница 132: ...110 The Advanced Model 4C3 User s Guide Notes ...
Страница 136: ...114 The Advanced Model 4C3 User s Guide Date Problem Symptom Action ...