Alarm Display Message
Is the
pump
infusing?
Alarm
Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
USING EXT BATT
Y
V
This is a normal response
when:
•
The pump has been
powered on with
external batteries as a
power source;
or
•
The pump has switched
to external battery
power due to a loss of
AC mains;
or
•
The pump has switched
to external battery
power while running
on internal batteries.
When powering on the
pump:
•
Press [
YES/ENTER
] to
confirm external battery
power source.
When switching from AC
mains:
•
Press [
SILENCE
] to
clear the alarm.
Alert Messages
Alert Display Message
Is the pump
infusing?
Alarm
Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
ALMOST EMPTY
Y
A/V
Delivery will complete
in 30 minutes or less.
For bolus only
program, delivery will
complete during the
next bolus delivery.
•
Press [
SILENCE
] to
quiet the alarm for 10
minutes.
Note: This alert is only
available with Pain
Management protocols.
94
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Troubleshooting
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...