Alert Display Message
Is the pump
infusing?
Alarm
Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
CHECK PRINTER
Y
A/V
•
The printer has not
responded to
commands from the
pump.
•
The printer is not
properly connected
to the pump.
1. Press [
SILENCE
] to
quiet the alarm for 2
minutes.
2. Ensure the printer is
properly connected to
the pump by checking
the following:
Þ
Printer is powered
on and in ONLINE
mode.
Þ
Cable is plugged in
to both the pump
and the printer.
Þ
Cable part number
and printer settings
are correct.
3. If the problem
continues, select
STOP
from the Print
Menu to cancel all print
jobs.
EMPTY CONTAINER
•
The pump does not
infuse unless Auto
KVO is delivering in a
mL/hr Only or
Weight Dosed
programs.
N
A/V
The pump has
delivered the
programmed container
size or the pump has
delivered the VTBI
programmed AND Auto
KVO is delivering.
1. Press [
SILENCE
] to
quiet the alarm for 2
minutes.
2. Press [
STOP
] to place
the pump in the
STOP
mode and clear the
alarm.
3. Use the
NEW
CONTAINER
function
to repeat the current
program. Refer to
Starting a New Container
in the “Basic
Operation” section.
Troubleshooting
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
95
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...