Abbott GemStar
™
Intermittent Programming Worksheet
GIVE:
Parameters
Selections and/or Values to be Entered
Programmable Ranges
Therapy
Intermittent
N/A
Dose Amount
0.1 - 9600 mL
Infusion Time Per Dose
hrs:min
1 min - 24 hrs
Dose Frequency
hrs:min
(Infusion time per dose + 1 min)
to 96 hrs
Number of Doses Per
Container
1 - 999
KVO
(Keep Vein Open)
q
No KVO Rate: 0.0
q
Yes KVO Rate:
0.1 -5.0 mL/hr
Container Size*
0.1 - 9999 mL
Air Sensitivity
q
On
Alarms at approx. 0.5 mL of air
q
2 mL Alarms at approx. 2 mL of air
q
Off
No alarm
N/A
Call Back Alarm
q
No
q
Yes
Alarm:
hrs:min before start
1 min - time between doses
Start Time
q
Start immediately after programming
q
Start time:
Specify AM or PM if not 24-hour clock.
12:00 AM - 11:59 PM or
00:00 - 23:59
* The pump automatically calculates the minimum container size required to deliver the
number of programmed doses and KVO between doses. If KVO is desired before a delayed
start, you must increase the container size to accommodate the desired KVO volume.
54
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Intermittent
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...