250 mL Lockbox (List 13024-01)
1
Open the lockbox (A).
2
Flip the top door up (B).
3
Align the pump with the lockbox (C).
4
Slide the pump down into the lockbox until fully seated.
5
Flip the top door down.
6
Connect the pump set to the bag and prime.
7
Place the flexible container inside the lockbox with the
administrative port down (D).
8
Hang the bag on the hook inside the lockbox (E).
9
Insert the cassette into the pump (F).
10
Ensure the tubing will not be pinched when closing the
lockbox.
11
Close and lock the lockbox.
12
Remove the key.
Removing the Lockbox
1
Remove the lockbox from the pole.
2
Unlock the lockbox.
3
Open the lockbox and flip the top door up.
4
Press the cassette release button to remove the cassette.
5
Lift the pump out of the lockbox.
Optional System Components
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
113
E
D
F
C
A
B
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...