Changing an Intermittent Program
d
Note: You cannot change a program when the pump is in
FULL
or
NEW CONTAINER
Lock.
Can be changed at
any time
Can be changed if no dose is
in progress
Cannot be changed after
[
START
] is pressed
• Dose parameters
• KVO rate
• Container size
• Air sensitivity
• Call back alarm
• Next dose start time
• Current time
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select [
4
]
CHANGE PROGRAM
.
2
As each program parameter displays, make the desired changes or press
[
YES/ENTER
] to accept the current setting.
3
Review the program.
4
Press [
START
] to begin delivery.