Troubleshooting
This section contains information on audible and visual alarms that may occur with the
Abbott GemStar Pump.
CAUTION:
If the pump does not perform as stated in this manual, stop using it
IMMEDIATELY.
For consultation and technical support contact your local Abbott Laboratories sales
office.
f
Tip: Press [
HELP
] for additional information during programming and alarm
conditions.
Alarm Messages
Alarm Display Message
Is the
pump
infusing?
Alarm
Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
AIR IN LINE
•
Pump automatically
enters
STOP
mode.
N
A/V
The specified volume of air
has been detected in the
line distal to the cassette
while infusing.
1. Press [
SILENCE
] to
quiet the alarm for 1
minute.
2. Press [
STOP
] to clear
the alarm.
3.
Disconnect the patient
from the pump set
.
4. Press and hold
[
PURGE
].
Note: Purge rate is 250
mL/hr. The pump purges
for up to 2 minutes each
time you press and hold
[
PURGE
]
.
5. Reconnect the pump set
to the patient-access
device.
6. Press [
START
] to begin
delivery.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
89
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...