Alarm Display Message
Is the
pump
infusing?
Alarm
Cause/
Condition
Remedy
Y = Yes
N = No
A = Audible alarm
V = Visual alarm
DIST. OCCLUSION
N
A/V
The pump has detected
sustained distal pressure
over the user-selected
limit.
To avoid a bolus infusion
after an occlusion,refer to
page 126.
1. Press [
SILENCE
] to
quiet the alarm for 1
minute.
2. Press [
STOP
] to place
the pump in the stop
mode.
3. Check for sources of
distal occlusion:
Þ
Kinked tubing below
the pump
Þ
Clamped patient-
access device
Þ
Clogged I.V. filter
Þ
Other obstructions
4. If the problem persists,
remove and reinstall
the cassette. Refer to
Loading the Cassette
in
the “System Setup”
section.
5. Press [
START
] to begin
delivery.
Troubleshooting
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
91
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...