Abbott GemStar
™
Variable Time Programming Worksheet
GIVE:
Parameter
Selections and/or Values to be
Entered
Programmable
Ranges
Select Phase Type
q
Percentages
q
Rates
q
Doses
0.1 - 100%*
0.1 - 400 mL/hr
0.1 - 9600 mL
Unit of Measure
q
Concentration:
mg/mL
q
Concentration:
mcg/mL
q
mL
0.1 - 1000 mg/mL
1 - 1000 mcg/mL
Current Time
Confirm or correct current time.
12:00 AM - 11:59 PM
or 00:00 - 23:59
Base Rate
q
No
q
Yes Start time:
Stop time:
Rate:
12:00 AM - 11:59 PM
or 00:00 - 23:59
0.1 - 400 mL/hr
Total Daily Amount for All Phases: __________ (required only when programming in percentages)
———- SEE ATTACHED WORKSHEET———-
Bolus Dose
0.1 mL - 25.0 mL
0.1 mg - 2500 mg
1 mcg - 25000 mcg
Bolus Lockout
minutes
5 - 999 min
KVO
(Keep Vein Open)
q
No KVO Rate: 0.0
q
Yes KVO Rate:
0.0 - 5.0 mL/hr
Container Size
(Total Daily Amount X Number of Days
of Therapy)
0.1 - 99999 mg
1 - 9999000 mcg
0.1 - 9999 mL
Air Sensitivity
q
On
Alarms at approx. 0.5 mL of air
q
2 mL Alarms at approx. 2 mL of air
q
Off
No alarm
N/A
*Sum of all phases must equal 100% when programmed in percentages.
Variable Time
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
67
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...