Change Pump Settings
1
Select
PUMP SETTINGS
from the Options Menu.
2
Select the number key that corresponds to the
desired pump setting.
3
Follow the display prompts. Refer to the
appropriate steps of the desired pump setting for
more information.
Select this item from
the Pump Settings
Menu:
To do this function:
Comments:
AIR SENSOR
Change the air sensor
sensitivity.
The air alarm sounds when the pump detects
an air bubble of approximately 0.5 mL when
ON; or at approximately 2 mL when 2 mL is
selected. The air alarm does
not
sound if OFF
is selected.
The air alarm may also be set during
programming if the current therapy allows it.
Removing power has no effect on the air alarm
setting.
OCCL SENSOR
Change the distal occlusion
sensor sensitivity.
Default Setting:
High
The pump alarms when distal pressure is
greater than the selected occlusion sensor
limit.
SOUND LEVEL
Adjust the audible alarm
level.
Default Setting:
Maximum
Options Menu
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
75
1 REVIEW PROGRAM
2 HISTORIES
3 KEYPAD LOCK
4 PUMP SETTINGS
5 SET CLOCK
6 PRINT
7 SPEED PROTOCOL
8 NEXT DOSE
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...