manualshive.com logo in svg
background image

ETCO

2

The ResQCPR™ System includes the following two components: 

1. ResQPUMP

®

 ACD-CPR Device

2. ResQPOD

®

 ITD 16

 

ResQCPR System Component:  

ResQPUMP ACD-CPR Device

 

DEVICE DESCRIPTION

The ResQPUMP ACD-CPR Device (hereafter  

referred to as ResQPUMP) (Figure 1) is a  

multi-use, hand-held device that includes a  

suction cup for attachment to the patient’s  

chest (with clothing removed), and a handle  

for the rescuer to hold onto. 

The ResQPUMP 

enables the rescuer to perform active compression- 

decompression cardiopulmonary resuscitation (ACD-CPR)

, which differs from  

standard CPR. During ACD-CPR, the chest is actively re-expanded (decompressed) after each  

compression; with standard CPR, the chest re-expands passively. The ResQPUMP design allows  

the operator to use the same body position and compression technique as in standard CPR.  

Active chest decompression is achieved when the rescuer maintains a firm grip on the ResQPUMP, 

bends at the waist and pulls his or her body weight upwards after compression. The suction cup 

sticks to the chest and transfers the lifting force to the middle of the ribcage. Compression force 

is transferred to the chest as in standard CPR via the device’s piston. The handle includes a force 

gauge that displays the forces exerted during both chest compression and decompression (chest 

wall recoil). The ResQPUMP has a battery-powered metronome integrated into the handle to guide 

the rescuer in the appropriate compression/decompression rate. The metronome emits two-tone 

signals of the same duration, one low and one high pitch tone. The signal (set at 80/minute)  

guides the rescuer to compress and decompress at the appropriate rate and for equal amounts of 

time (50% duty cycle).

 

If suction difficulties occur, adjust the angle of the ResQPUMP on the chest to obtain an adequate 

seal. It may be necessary to shave hair from the middle of the chest to achieve good suction. 

NOTE: If suction difficulties persist, the ResQPUMP can still be used for compressions without  

causing additional harm to the patient, as long as it does not distract from CPR quality. 

ResQCPR System Component: ResQPOD ITD 16 

Instructions For Use

All users must read and fully understand these 

ResQCPR System Instructions for Use (IFUs) 

before using the ResQCPR System  

or any of its components. The ResQCPR™  

System includes the following components: 

1. ResQPUMP

®

 ACD-CPR Device

2. ResQPOD

®

 ITD 16

 

USER INFORMATION: The use of other  

medical equipment or drugs in conjunction  

with cardiopulmonary resuscitation (CPR) may 

impact the effect of compressions/decompres-

sions. Consult the instructions for use for other 

equipment or drugs to assure that they are  

used appropriately in conjunction with active  

compression decompression CPR (ACD-CPR) 

with an impedance threshold device (ITD) -  

the ResQCPR System. 

 

REQUIRED SKILLS: The ResQCPR System  

should be used only by persons with  

appropriate training, such as ambulance and 

rescue personnel, or other medical staff who 

have already completed a conventional CPR 

course (e.g. American Heart Association,  

American Red Cross, or equivalent), and who 

have been trained on use of the ResQCPR 

System. 

  

 

RX ONLY 

Federal Law (USA) restricts this 

device system to sale by or on 

the order of a licensed medical 

practitioner.

 

CONTRAINDICATIONS

None known.

WARNINGS

Improper use of the ResQCPR System could cause serious injury to the patient and 

ineffective chest compressions/decompressions. The ResQCPR System should only  

be used by personnel who have been trained in its use.

 

Improper positioning of the ResQPUMP suction cup may result in possible  

injury to the rib cage and/or internal organs, and may also result in suboptimal  

circulation during ACD-CPR.

 

Do not use the ResQPUMP if the patient’s chest is not large enough for the ResQPUMP 

suction cup to provide adequate compressions/decompressions during use. 

 

Moisture, gels, or other lubricating materials on the patient’s chest should be removed 

before applying the ResQPUMP. Failure to do so may result in sliding  

of the suction cup on the chest, ineffective chest compressions/decompressions, and 

possible injury to the rib cage or internal organs. 

 

The ResQPUMP should not be used in patients who have had a recent  

sternotomy. Use of the ResQPUMP in patients with a recent sternotomy  

(within the past 6 months) has not been evaluated, but this may potentially cause 

serious injury. 

The safety and effectiveness of CPR using the ResQCPR System have not been  

assured when used to treat cardiac arrest in patients with drug/medication  

overdose etiology.

If the patient has a return of spontaneous circulation (ROSC) (e.g. palpable pulse) 

during the resuscitation efforts, the ResQPOD should be immediately  

removed from the airway circuit and use of the ResQPUMP should be  

discontinued. Failure to do so may cause shortness of breath, difficulty breathing and 

potential pulmonary edema if the patient begins to breathe spontaneously.

The ResQCPR System has not been studied in the setting of in-hospital cardiac arrest; 

therefore, the safety and effectiveness of the device in this setting are unknown.  

Clinical outcomes when using the ResQCPR System in the setting of an in-hospital 

cardiac arrest. Therefore, the safety may be different from the outcomes observed in 

the clinical trials of out-of-hospital cardiac arrest.

Safety and effectiveness of the ResQCPR System in pregnant women and  

children under the age of 18 have not been studied in clinical trials.

Safety and effectiveness of the ResQCPR System in the setting of traumatic  

injury (wounds resulting from sudden physical injury) have not been established.

PRECAUTIONS

Figure 1:

 ResQPUMP ACD-CPR Device

Metronome

Force Gauge

Handle

Suction Cup

Figure 2:

 ResQPOD ITD 16

On/Off Switch for 

Timing Assist Lights

Inspiratory  

Impedance  

Regulator

Timing Assist Lights

Safety Check 

Valve

INSPECTION PRIOR TO USE

Carefully inspect the ResQPOD in its package 

before opening. Do not use if package is 

opened or damaged, or if any defects are 

noted.

HOW SUPPLIED

The ResQPOD ITD 16 is non-sterile and  

intended for single patient use only. Dispose  

of properly when use is completed.

DEVICE DESCRIPTION

The ResQPOD ITD 16 (hereafter referred to 

as ResQPOD) (Figure 2) is a non-sterile, single 

patient use, disposable impedance threshold device (ITD) that 

limits passive air entry into 

the lungs during the chest wall recoil (decompression) phase of CPR

, thereby 

reducing intrathoracic pressure when rescuers are not providing a breath. Lowered intrathoracic 

pressure results in greater venous return (preload) which, in turn, results in greater cardiac output 

on the subsequent compression. It is inserted into the airway circuit between the patient and  

the ventilation source, and can be used with either a facemask or advanced airway (e.g.  

endotracheal [ET] tube). The ResQPOD may be used with standard ventilation sources (e.g.  

bag-valve or demand-valve resuscitators, rescuer’s mouth, automated ventilator). It does not  

restrict the patient’s ability to exhale, nor the rescuer’s ability to ventilate. The ResQPOD allows 

the rescuer to provide periodic positive pressure ventilation while impeding passive inspiratory 

gas exchange during the chest recoil phase. The ResQPOD ITD 16 includes a safety check  

valve that allows inspiration at -16 cmH

2

O. The check valve is a design safety feature in the 

event that the patient begins to breathe independently while the device is in place within the 

airway circuit. Timing assist lights provide guidance to the rescuer on the proper ventilation rate 

for a patient with an advanced airway. 

PERFORMING CPR USING THE RESQCPR SYSTEM

Before beginning CPR with the ResQCPR System  

(ACD-CPR with an ITD; Figure 3), always assess  

the patient according to local standards to assure  

there are no signs of circulation (e.g. consciousness,  

breathing, coughing, movement or pulse). Begin  

performing CPR with the ResQCPR System as soon  

as possible but do not delay manual chest  

compressions while preparing the ResQCPR devices. 

PERFORMING CPR USING THE RESQCPR  

SYSTEM IN ADULTS

1.  Assure that patient is pulseless and that  

resuscitation is indicated.

2.  Place ResQPUMP; turn on metronome and begin  

performing compressions to appropriate depth  

(e.g. 2” or 5 cm) depth at rate of 80/min (see  

additional detailed instructions on the use of the ResQPUMP below).

3.  Attach ResQPOD to facemask as soon as chest compressions begin; use a  

2-handed technique to maintain a tight facemask seal and airway position  

(see additional detailed instructions on the use of the ResQPOD below).

4.  After 30 compressions, pause and administer two ventilations over one  

second duration each until the chest rises.

5.  Continue to provide a 30:2 compression to ventilation ratio until pulse returns or  

advanced airway is placed. Rotate ACD-CPR duties every two minutes to avoid fatigue.

6. Once an advanced airway (e.g. ET tube, supraglottic airway) is placed:

     • Confirm tube placement and secure with commercial tube restraint.

     •  Move the ResQPOD to the airway and turn on the timing assist lights.

     •  Provide asynchronous ventilations; ventilate once (over one second until chest  

rises) each time light flashes (10/min).

     •  Perform continuous chest compressions at 80/min. Do not pause compressions  

for ventilations. 

     •  Rotate ACD-CPR duties every two minutes to avoid fatigue.
7.  If the patient has a return of spontaneous circulation (ROSC) the ResQPOD should  

be immediately removed from the airway circuit and use of the ResQPUMP should be  

discontinued. If the patient re-arrests, resume use of the ResQCPR System immediately.

 

Figure 3:

 Use of the ResQCPR System

NOTE: 

Signs and symptoms of improved 

cerebral blood flow (e.g. eye opening,  

gagging, spontaneous breathing, limb or  

body movement) have been reported in  

patients without a spontaneous pulse but  

who are undergoing resuscitation with the 

ResQCPR System. If these are noted, check 

quickly to see if a spontaneous pulse has  

returned. If the patient remains in cardiac 

arrest, continue resuscitation with the  

ResQCPR System and contact your medical 

control authority or local protocol for  

guidance on managing these signs and  

symptoms in an arrested patient. If a  

spontaneous pulse has returned, discontinue  

the ResQCPR System and support ventilations 

as indicated. 

THE RESQCPR TEAM

It is highly recommended that rescuers work  

in teams of 3 - 4 people. This enables one  

person to perform ACD-CPR, one person  

to hold the facemask in place with the  

ResQPOD attached, and a third person to  

manage the defibrillator and observe the  

patient for signs of ROSC. In this team  

scenario, the person compressing the chest  

can stop after 30 compressions to provide  

two positive pressure breaths (<5 seconds),  

before resuming chest compressions. With 

more personnel, two rescuers can manage  

the airway. 

ENSURING HIGH QUALITY CPR WITH 

THE RESQCPR SYSTEM

•  Check the ResQPUMP’s force gauge at  

regular intervals to ensure that the  

appropriate forces needed to compress  

and decompress are being delivered. 

•  Use the ResQPUMP’s metronome as a  

guide for compressing and decompressing 

at the rate of 80/min. NOTE: This rate is 

slightly slower than the rate recommended  

for standard CPR in order to allow sufficient 

time for blood return to the chest, and to 

reduce rescuer fatigue.

•  Rotate ACD-CPR duties every two minutes  

to avoid fatigue.

•  Once the patient has an advanced airway 

placed, use the ResQPOD’s timing assist  

lights to guide the proper ventilation rate. 

Ventilate over one second duration and do 

not hyperventilate!

• Avoid unnecessary interruptions.

DETAILED INSTRUCTIONS RELATED TO THE USE OF THE RESQPUMP

Proper use of the ResQPUMP is shown in 

Figures 4 through 9

.

RESQPUMP POSITIONING

The ResQPUMP’s compression point is the same as for standard 

manual CPR (Figure 4). Position the suction cup in the middle of the 

sternum between the nipples (mid-nipple line). Make sure that the 

edge of the suction cup does not extend below the xiphoid process, 

as this could result in inadequate suction and/or rib injury.

 

RESCUER POSITION

Kneel close to the patient’s side. For optimal position, shorter 

rescuers may find it beneficial to be slightly elevated by kneeling on 

padding. If the patient is in bed (with hard surface under torso), it 

will be necessary to kneel next to the patient or stand on a platform 

of sufficient height. Grasp the ResQPUMP’s handle with both hands, 

placing the heels of the hands near the gauge with wrists slightly 

bent (Figure 5).

 

COMPRESSIONS WITH THE RESQPUMP

Compress the chest to the recommended depth (e.g. 2” or 5 cm), 

observe the force required to achieve that depth, and then use that 

force target as a guide. The amount of force required will vary 

according to how compliant the chest is. Compress with shoulders 

directly over the sternum, with arms outstretched and elbows locked 

(Figure 6). Use the large thigh muscles to compress, bending at the 

waist. Compress at a rate of 80/min and use the metronome to 

guide the compression rate. Start and stop the metronome by  

pressing the red button on the force gauge (Figure 7). Compress the 

chest on one tone and lift on the other tone.  

Note: The compression rate using the ResQPUMP  

(80 compressions per minute) is different than the 

current AHA-recommended rate of 100-120  

compressions per minute for standard CPR.

 

COMPRESSION FORCE

The red arrow tip indicates the force being applied (Figure 7).  

The approximate amount of force required to compress the  

chest two inches (5 cm) is as follows:

• 30 kg (

 65 lb) of force: soft/supple chest

•  30 - 40 kg (

 65 - 90 lb) of force: chest of medium/average 

compliance (Figure 7)

• 50 kg (

 110 lb) of force: stiff/rigid chest

Once it has been determined how much force is required to  

compress the chest to the appropriate depth, use that amount of 

force as a guide for continued compressions.

 

DECOMPRESSIONS WITH THE RESQPUMP

To provide active decompression, use the large muscles in the thighs 

to lift, bending at the waist (Figure 8).

 

DECOMPRESSION FORCE

Decompress (lift) the chest until the tip of the red arrow on the force 

Figure 4

Figure 5

gauge registers -10 kg (

 –20 lb) of force (Figure 9). This amount of upward  

force must be exerted to fully achieve the benefits of active decompression.  

Closely monitor the force gauge and suction cup seal during use. If the suction 

cup dislodges, reposition it with the next compression; then, on the next  

decompression, lift until just before the suction cup releases but do not exceed  

-10 kg (

 -20 lb) of lift. Use a 50% duty cycle, spending equal time compressing 

and decompressing.

 

SUCTION CUP REMOVAL 

Lift up an edge of the suction cup lip to release the vacuum under the cup.  

This will free the cup from the patient’s chest.

Figure 6

Figure 7

Decompression

Compression

30

0

kg

40

15
10

50

80/min

LIFT

 to 10 kgs

Use enough

force to

COMPRESS 

to 2” (5 cm) depth

Figure 8

Figure 9

TROUBLESHOOTING

1.  If suction difficulties occur, adjust the angle of the ResQPUMP on the chest to obtain an  

adequate seal. It may be necessary to shave hair from the middle of the chest to achieve  

good suction. NOTE: If suction difficulties persist, the ResQPUMP can still be used for  

compressions (with the metronome disabled) without causing additional harm to the  

patient, as long as it does not distract from CPR quality. 

2.  Rib fractures can occur with any method of CPR, even if performed correctly. If it appears  

that rib fractures have occurred, check to make sure the suction cup is properly positioned  

and that compression depth is appropriate. The occurrence of rib fractures is not sufficient  

reason to discontinue ACD-CPR. 

3.  If there are questions about whether the ResQPUMP is functioning properly, consider  

discontinuing its use and perform standard manual CPR instead. 

DETAILED INSTRUCTIONS RELATED TO THE USE OF THE RESQPOD ON  

A FACEMASK

1.  It is important to insert the ResQPOD into the ventilation circuit as soon as 

chest compressions begin. In most cases this will involve placement on a  

facemask; however, never delay the initiation of chest compressions while 

waiting to place the ResQPOD. 

2.  The ResQPOD may be used on patients with or without an oral or nasal 

airway. 

3. Attach the ResQPOD to the facemask. (Figure 10) 
4. 

 Obtaining and maintaining a tight facemask seal throughout 
both chest compressions and ventilations is critical.

 To achieve 

this, spread out the cushion of the mask. (Figure 11) 

5.  Use a two-handed technique to maintain proper airway positioning and 

obtain a tight facemask seal. Place the facemask onto the patient,  

covering the nose and mouth. Obtain a tight facemask seal by either:

     A. Using the thumb and base of the palm (Figure 12); or

     B. Forming a “C” with thumb and index finger.
6.  Place the remaining fingers on the bony part of the lower jaw and lift the 

lower jaw to the facemask. Do not push the facemask into the face to try  

and obtain a seal. 

7.  Tilt the head back and continue to lift the lower jaw to the mask.  

(Figure 13)

8. Attach the ventilation source to the top of the ResQPOD. 
9.  DO NOT turn on timing assist lights. If rescue personnel elect to  

perform continuous chest compressions and ventilate asynchronously with a 

facemask, then the lights may be turned on and the rescuer should provide  

a positive pressure ventilation on the upstroke of ACD-CPR every time the 

light flashes.

10.  Administer ResQCPR using the proper compression to ventilation ratio  

(e.g. 30:2) and the proper ventilation duration (e.g. one second). 

11.  If there is only one rescuer available to manage the airway, then the  

rescuer who is performing chest compressions can reach over and  

provide ventilations during the pause in chest compressions, while  

the rescuer at the airway maintains a two-handed facemask seal.  

(Figure 13) 

12.  The ResQPOD is disposable and intended for single patient use. Cross 

contamination may occur if the device is used on multiple patients.

 

1.  Insert the advanced airway and confirm tube placement. Secure the tube 

with a commercial tube restraint device. The use of tape to secure the tube 

is not recommended. 

2.  Attach the ResQPOD to the top of the airway and attach the ventilation 

source to the top of the ResQPOD (Figure 14); avoid interrupting CPR to  

do this.

3.  Turn on the timing assist lights. To activate, slide the ON/OFF switch to  

the ON position. 

4.  Perform ResQCPR with continuous chest compressions (no pauses for  

ventilations). Ventilate asynchronously when the timing lights flash (10/min) 

and over proper duration (e.g. one second) until the chest rises.

5.  If an end-tidal carbon dioxide detector is used, place it in the airway circuit 

between the ResQPOD and the ventilation source. (Figure 15) 

6.  If there is desire to administer medications via the advanced airway,  

remove the ResQPOD and administer medications directly into the tube,  

then re-connect ResQPOD and resume use. 

ADDITIONAL INSTRUCTIONS:

1.  If there is a return of spontaneous circulation (e.g. palpable pulse) and chest 

compressions are no longer indicated, immediately remove the ResQPOD 

from the airway circuit. Failure to do so may cause shortness of breath and 

difficulty breathing. Support respirations as indicated.

2.  Immediately replace the ResQPOD in the ventilation circuit if the patient 

re-arrests and chest compressions are again indicated. 

3.  If vomit or secretions enter the ResQPOD, remove the ResQPOD from the 

facemask or airway adjunct and use the ventilation source to clear the  

material from the ResQPOD. Suction the patient as needed, then re-attach 

the ResQPOD and resume use. Discontinue use of the  

ResQPOD if it cannot be cleared or if positive pressure ventilation is  

compromised in any way with the device in place.

Figure 10

Figure 11

Figure 12

Figure 13

DETAILED INSTRUCTIONS RELATED TO USING RESQPOD ON AN ADVANCED  

AIRWAY (E.G. ET TUBE):

Figure 14

Figure 15

FOLLOWING EACH USE OF THE RESQCPR SYSTEM 

The ResQPOD is intended for single patient use only and should be discarded after use. Failure to 

do so may cause cross contamination between patients. The ResQPUMP should be cleaned and 

disinfected after every use. 

 

RESQPUMP CLEANING

To clean the handle, wipe with damp cloth and mild detergent. The suction cup may be replaced 

with a new suction cup, or cleaned. To clean the suction cup, wash it with a mild detergent and 

rinse with tap water. 

 

 Never immerse the handle in water or autoclave to clean. Doing so may cause  

permanent damage. 

 

RESQPUMP CHEMICAL DISINFECTION

The handle and suction cup may be chemically disinfected after washing. Wipe the cup and  

handle with a bleach solution (5% chlorine, minimum) or Cavicide

®

 (follow manufacturer  

instructions for wetting times). Wipe the handle with a dampened cloth to remove chemical  

residue. Do NOT immerse the handle. The cup may be rinsed with water. Wipe with a clean  

dry cloth and allow to air dry. 

 

 Hazards during disinfection. Always wear protective clothing during disinfection  

of the ResQPUMP. Follow the handling instructions from the manufacturer of  

the disinfectant. 

 

NOTE: The cleaning procedure is sufficient after ‘normal’ soiling. If there are bodily fluids on the 

ResQPUMP or if an infectious patient has been treated, the ResQPUMP should also be disinfected 

as described above and the suction cup should be discarded and replaced.

RESQPUMP FUNCTION TESTING

Before placing the ResQPUMP into service and following each use, the following functional tests 

should be performed:
1.  Inspect the handle and suction cup for visible damage. Do not use the ResQPUMP if there is 

obvious damage to the suction cup or handle. NOTE: Replacement suction cups are available 

from the manufacturer. 

2.  Compress the ResQPUMP against a smooth hard surface with approximately 50 kg of force, 

using the force gauge on the ResQPUMP as a guide. Observe for an increasing gauge  

reading. 

3.  Pull up on the handle with approximately 10 to 15 kg of force, using the decompression  

force gauge as a guide. Observe for a decreasing gauge reading and check for proper  

suction. The gauge should move smoothly within the compression and decompression ranges. 

4.  Ensure that the force gauge reads zero (Figure 16) when no force is applied. If it does NOT 

read zero, see instructions for force gauge readjustment below.

5.  Assess the metronome’s battery level by pressing on the metronome button for more  

than three seconds. If the battery is okay, first, a long high-note beep will be heard,  

followed by three short beeps. If one long low-note beep is heard, or if no beep is heard,  

the device should be replaced as well. 

INDICATIONS FOR USE

The ResQCPR System is intended for use as  

a CPR adjunct to improve the likelihood of 

survival in adult patients with non-traumatic 

cardiac arrest.

1

2

3

4

5

6

7

8

9

Отзывы: