Changing a Continuous Program
d
Note: You cannot change a program when the pump is in
FULL
or
NEW CONTAINER
Lock.
Can be changed at any time
Cannot be changed
• Delivery rate
• VTBI
• KVO rate
• Container size
• Air sensitivity
• Piggybacking
• Unit of measure
• Concentration
To change a program:
1
From the
STOP
mode, press [
CHANGE
] and select [
4
]
CHANGE PROGRAM
.
2
As each program parameter displays, make the desired changes or press
[
YES/ENTER
] to accept the current setting.
3
Review the program.
4
Press [
START
] to begin delivery.
Continuous
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
63