8
Push down on the flow stop
to ensure the cassette
is in the closed position.
9
Insert the cassette into the cassette pocket. Refer to “Loading the Cassette” on page
26.
d
Note: Extension set with anti-siphon valve must be attached before patient-use.
10
Connect the extension set to the pump set (refer to the instructions that come with
the pump set).
11
Press [
PURGE
].
PRIME THE SET?
displays. Press [
YES/ENTER
].
12
Press and hold [
PURGE
].
d
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press
and hold
[
PURGE
]
.
Continue using the PURGE key until fluid fills the tubing. Ensure all air is removed
from the container, cassette, and tubing.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of
the pump set. If flow is observed, DO NOT use the pump set.
13
If extension set has a Y-extension, prime it if needed.
14
Connect the pump set to the patient-access device.
System Setup
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
25
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...