For example, in the following diagram, an intermittent dose was started at 2:00. The
dose was interrupted during the current dose cycle and was started again before the
end of the current dose cycle.
Dose Delivered -
Dose Interrupted -
6
If an Intermittent dose is interrupted and the pump is placed in the
RUN
mode after
the off cycle of the current dose, the next dose begins delivery and all subsequent
doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was stared at 2:00. It
was interrupted during the current delivery cycle at 2:30, but was not started again
until after the current delivery cycle at 6:45:
Dose Delivered -
Dose Interrupted -
6
When the dose is complete, the start time for the Next Dose displays. To change the
start time refer to the “Options Menu” section.
6
Any missed Intermittent doses, except the first dose, are added to the end of the
delivery schedule.
52
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Intermittent
12 1
2
3
4
5
6
7
8
9
10 11 12
12 1
2
3
4
5
6
7
8
9
10 11 12
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...