Abbott GemStar
™
TPN Programming Worksheet
GIVE:
Parameters
Selections and/or Values to be
Entered
Programmable Ranges
Therapy
TPN
N/A
Delivery Mode
q
Continuous
q
Cont Taper Down
q
Cont Taper Up
q
Cont Taper Up and Down
N/A
VTBI
(Volume To Be Infused)
mL
1 - 9600 mL*
Taper Up
q
No
q
Yes
hrs:min
1 min - 3 hrs
Taper Down
q
No
q
Yes
hrs:min
1 min - 3 hrs**
Total Time to Infuse
hrs:min
1 min - 24 hrs
KVO
(Keep Vein Open)
q
No KVO Rate: 0.0
q
Yes KVO Rate:
1.0 - 5.0 mL/hr
Container Size
(VTBI + KVO)
mL
9999 mL max
Air Sensitivity
q
On
Alarms at approx. 0.5 mL of air
q
2 mL Alarms at approx. 2 mL of air
q
Off
No alarm
N/A
* Maximum VTBI is 9600 mL minus any taper amount.
** Programmable ranges for taper down are 1 minute to 3 hours OR 2 times the remaining volume
divided by the continuous delivery rate, whichever is less.
Total Parenteral Nutrition (TPN)
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
57
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...