Priming Pump Sets WITH a Drip Chamber
WARNING: Failure to use the anti-siphon valve may result in unrestricted
flow.
ALWAYS set the Air Sensitivity to either ON or 2mL when
using vented drip chambers.
1
Push down on the flow stop
to ensure the cassette
is in the open (priming) position.
2
Close the CAIR® (roller) clamp.
3
Insert the piercing pin into the container. Suspend the
container on an I.V. Pole.
4
Squeeze the drip chamber until filled halfway.
5
Invert the cassette approximately 45°.
6
Open the CAIR® (roller) clamp.
System Setup
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
21