Using the Abbott GemStar Docking Station
(List 13075-01)
Connecting the Pump
1
Mount the docking station to the I.V. pole and tighten the
knob.
2
Align the connectors on the bottom of the pump with the
connectors on the docking station. Slide the pump into the
docking station until it snaps into place (A).
3
Connect the docking station to AC mains power.
The docking station light illuminates yellow while the docking
station is charging. When the docking station is fully charged, the light illuminates
green.
Connecting the Bolus Cord and Printer Cable
WARNING: Disconnect the pump from the patient BEFORE connecting the
pump to a printer or computer.
1
Plug the bolus cord (B) into the bolus port on the rear,
right side of the docking station.
2
Plug the printer cable (C) into the printer port on the rear,
right side of the docking station.
Optional System Components
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
119
B
C
A
Содержание GEMSTAR
Страница 7: ...vi Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 21: ...14 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 35: ...28 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 65: ...58 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 91: ...84 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 95: ...88 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 105: ...98 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 109: ...102 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 143: ...136 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...
Страница 145: ...138 Abbott GemStar Manual 430 600095 002 Rev 8 02 Notes ...