Loading the Cassette
1
Align the cassette as shown.
2
Insert the cassette into the
cassette pocket.
3
Push the cassette into the
pocket until it is firmly seated.
When properly installed, the
cassette cannot be removed
from the pocket unless the
cassette release button is
pushed.
CAUTION:
Ensure the cassette is
properly installed.
4
Align the tubing in the tubing
channel.
CAUTION:
Before connecting to
the patient-access
device, ensure no
fluid flows at the
distal end of the
pump set while the
pump is in the
STOP
mode. If flow is
observed, DO NOT use
the pump set.
5
Connect the pump set to the patient-access device.
26
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)