background image

Titan S:

 

Tube Assembly, Gasket, Retainer Ring, (2/3-Line) ..........................256454

Tube Assembly, Gasket (4-Line) .............................................................256455

Instruction Manual

Titan

®

 Blis-Sonic Scaler SW/S

Replacement Assemblies and Spare Kits

WARNING:

 Never remove scaler housing with tip assembled to the scaler.

1. Unthread the front of the scaler housing to expose the front suspension carrier.

2. Remove the two old O-rings and install the replacement O-rings, 

StarDental

 Part #256718.

3. Re thread front of scaler housing to original position.

Titan S/SW Rear Seal & Swivel O-Rings

O-Rings & Rear Seal Replacement 

Swivel O-Rings:

  Periodically replace the swivel 

O-rings, 

StarDental

®

 Part #259884.

Front Suspension O-Rings:

 After two years, front 

suspension O-rings relax resulting in play between 

shaft and handle. 

Swivel O-Rings

Front Suspension O-Rings   

Front of Scaler Housing

Seals wear with repeated use. Replace rear seals 

annually or as needed. On either the  

2/3-line swivel or Titan S 2/3-line Scaler replace 

the 2/3-line rear seal with 

StarDental

 Part #203103. 

On either the 4-line swivel or Titan S 4-line Scaler 

replace seal with 

StarDental

 Part #202897.

Rear Seal on 

the Swivel

Rear Seal on 

the 4-line

Titan S

2/3-line for Titan S/SW

1. Remove 2/3-line rear seal with an 

explorer.

2. Place new rear seal over drive air and 

water tubes.

3. Carefully push seal into place using a 

blunt instrument.

 Rear Seal 

 Titan S 2/3-line

 Titan SW

2/3-line

 Rear Seal 

 Swivel

Cleaning Wire 

CLEANING the Tips

Clean tips weekly by running cleaning wire, 

StarDental

 Part #256719, through entire tip starting at 

opening in front.

CLEANING Titan SW or S Water Tube

On a regular basis or if a water tube blockage occurs, 

remove the tip, and on the Titan SW also disconnect the 

swivel. Push the cleaning wire through the internal water 

tube using needle nose pliers at the front on both the Titan 

SW and S as shown. Push small lengths of 3 to 4 mm at a 

time until the wire emerges at the opposite end. Pull wire 

completely out.

Cleaning  Wire

Opening

Front - Titan K and SW 

StarDental®

 

Instruments

Reviews: