6
SEITE 6 - 148 x 210 mm - 3100021788/01 - SCHWARZ - 20-052 (kn)
Battery Power Level Display on Handpiece
Operation
Selection of Exposure Time
Select the exposure time by pressing the TIME button.
- The selected exposure time is indicated by the
4 green LEDs.
- Each time the button is briefly pressed, the setting
advances to the next (higher) value. All 4 green
LEDs will be turned on for a setting of 20 sec.
Pressing the button again will turn off all of the
LEDs and enable the continuous mode.
- The display advances through the available
settings if the button is kept depressed.
- While exposure is ongoing, the button for selection
of the exposure time is inactive.
Activating and Deactivating the Light
왘
Briefly press the START button; the light will turn on.
- The LEDs first indicate the set exposure time; all
4 LEDs light up for 20 sec. Every 5 sec, as the time
runs down, the LEDs will turn off one at a time;
at 15 sec remaining time, 3 LEDs will still be on,
at 10 sec remaining time 2 LEDs, etc.
- The LEDs do not come on at all in continuous mode;
an audible signal is emitted every 10 seconds.
왘
If desired, the light can be turned off by pressing the
START button again before the exposure time is over.
왘
Holding down the START button activates the tack-
cure mode: the device emits a single short light pulse
which enables the defined curing of Protemp
™
Crown
temporary restorations or a light-curing cement
excess (e.g., RelyX
™
Unicem) to enable easy removal.
Inserting and Removing the Light Guide from/into
the Handpiece
왘
Attach the light guide to the handpiece by inserting it
with a slight rotation until it clicks firmly into place
and there are no gaps between the neck of the light
guide and the handpiece (see picture).
LED showing
exposure times
{
START button
TIME button
0
1
2
3
4
5
6
7
8
9
10
Distance (mm)
Light intensity (%)
100
90
80
70
60
50
Status LED
Operating status
Handpiece without
Charger connected
charger
Steady green light
Handpiece ready
Charging has been
for operation
completed
Flashes green
––
Battery is being
charged
Steady red light
Low battery charge Problem during
charging
Flashes red
Battery fully
Problem in
discharged,
charging, battery is
exposure cycle will defective or cannot
be completed or,
be charged
if in continuous
mode, stopped
Exposure time options: 5, 10, 15, 20 sec, continuous
mode (120 sec), tack-cure mode (1 sec).
왘
See the instructions for use for the specific dental
material when selecting the exposure time.
왘
The indicated exposure times assume that the light
guide is held at the exact position over the material
being polymerized.
왘
If the distance between the light guide and the
restoration is increased, the exposure time must be
adjusted accordingly because the light intensity
weakens (see graph).