Maintenance and Care
The Elipar DeepCure-L device is maintenance-free.
No periodic maintenance is required. See the informa-
tion contained in this chapter to secure problem-free
operation.
Care of the Handpiece
왘
Use only the charger included with the product. The
use of other chargers may damage the battery cells
or result in inadequate charge.
Do not immerse the handpiece in water or incinerate.
Please also observe the chapter on “Safety”.
Reprocessing the Light Guide
The light guide is not sterile when delivered and must
be reprocessed before being used for the first time and
after each patient use.
Follow manufacturer’s instructions for preparation, tem-
peratures, contact time and rinsing off the cleaning
1
and disinfecting solution
2
exactly.
An enzymatic detergent should be used as the cleaning
solution.
Material Resistance
Make sure that the cleaning and disinfectant agents you
have chosen do not contain any of the following materi-
als:
– Organic, mineral, and oxidizing acids (minimum ac-
ceptable pH value 5.5)
– The pH value of the cleaning and disinfecting solu-
tion should be between pH 7 and 11
– Oxidation agents (e.g., hydrogen peroxide)
– Halogens (chlorine, iodine, bromide)
– Aromatic/halogenized hydrocarbons
Error
Cause
왘
Solution
The dental material
does not cure
completely.
The light guide cannot
be attached to the
handpiece.
왘
Clean the light guide and
the protecting glass in the
light guide mounting hole
(please refer to “Cleaning
the Light Guide”).
왘
Check to see that the
correct light guide has
been attached.
왘
The light guide is not
designed for use with the
Elipar DeepCure-L.
Error
Cause
왘
Solution
While the handpiece is
connected to the
charger, pressing the
START button will not
start the light emission.
The light emission
does not start when
the START button is
pressed; an error
signal sounds for
2 sec.
During light emission
in continuous mode,
an error signal sounds
for 2 sec, the emission
is stopped, and the
handpiece changes
to sleep mode.
The light intensity is
too low.
The handpiece is connected
to the charger. Light-curing
is prevented as a safety
precaution.
왘
Disconnect the charger
from the handpiece and
restart the light emission.
The information about inter-
mittent operation under the
header “Technical Data -
Handpiece” has not been
observed.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
왘
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
The information about inter-
mittent operation under the
header “Technical Data -
Handpiece” has not been
observed.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
왘
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
왘
Clean the light guide and
the protecting glass in the
light guide mounting hole
(please refer to “Cleaning
the Light Guide”).