54
9650-0804-01 Rev. A
14
Remove the NIBP module from the unit.
15
Tilt side panel to access the EtCO
2
connector and use an orange stick release
the tabs and remove the EtCO
2
connector.
16
Remove the EtCO2 module from the unit.
17
Remove three (3) screws from the outside
of the left side panel in the areas illustrated.
Note: During reassembly, torque screws to
6 in-lbs.
18
Remove four (4) screws from the inside top
of the left side panel in the areas illustrated
and carefully pull out side panel.
Note: During reassembly, torque screws to
6 in-lbs.
19
Remove ECG ZIF connector in the area
illustrated.
20
Remove the left side panel from the unit.
EtCo
2
connector
before removal
EtCo
2
connector
after removal
EtCO2 connector after removal