Functional Verification Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
23
Leads Test
Tools Needed
3-lead, 5-lead, and 12-lead cables
Test each cable separately.
Test Setup
Connect the lead wires appropriate to each test to the
Fluke Impulse 7000DP Analyzer
or
equivalent.
Do this...
Observe this...
Pass/Fail
1.
Press the power switch to turn on the
Propaq M.
No ECG LEAD FAULT message
displayed.
o
o
2.
Disconnect one lead from the
simulator.
The ECG LEAD FAULT message displays
within 3 seconds (if configured).
o
o
3.
Reconnect the lead. Repeat step 2
with the remaining leads.
Wait for ECG LEAD FAULT message to
clear from the display (if configured).
o
o
4.
Repeat steps 2 and 3 for 5 lead and 12
lead cables.
Note
: If heart rate alarm sounds, on the
Propaq M:
•
Press the Next quick access key
(
) key to access the second
set of quick access keys.
•
Press the Alarm
quick access key
(
).
•
Press the Alarm Suspend key
(
).
o
o
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...