52
9650-0804-01 Rev. A
3
Disconnect multi-wire IBP cable with ferrite
in the area illustrated.
4
Cut the tie wrap to the IBP cable in the area
shown with the orange stick.
Note: During reassembly, secure a new tie
wrap to the IBP cable.
5
Disconnect the USB flex cable from ZIF
connector in the area illustrated.
6
Lay front end of separated unit down with
opened side facing up.
7
Disconnect the black EtCO
2
tube
connector from the air vent.
8
Remove four (4) screws in the areas
illustrated to release the silver EtCO
2
module.
Note: During reassembly, torque screws to
6 in-lbs.
multi-wire IBP cable
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...