30
9650-0804-01 Rev. A
NIBP Test
Tools Needed
NIBP simulator, single or dual lumen hose, adult, NIBP cuff, rigid cylinder
Test Setup
Connect the Propaq M to the NIBP simulator using the single or dual lumen hose. Wrap the
NIBP cuff around a rigid cylinder, such as a glass jar.
Note:
Refer to Appendix B for information on simulators and settings that you may use to
assess the performance of the NIBP module.
Do This...
Observe This...
Pass/Fail
1.
On the Propaq M, ensure the patient mode
is set to Adult.
2.
Set the simulator to the 80/50 simulator
preset.
3.
Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
4.
Set the simulator to the 120/80 simulator
preset.
5.
Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
6.
Set the simulator to the 200/150 simulator
preset.
7.
Take a manual reading.
Confirm that the NIBP measurement
reported is /- 10 mmHg of
the expected simulator performance.
o
o
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...