C
HAPTER
1
M
AINTENANCE
T
ESTS
28
9650-0804-01 Rev. A
EtCO
2
Test
Tools Needed
CO
2
sampling line airway adapter
Test Setup
None
Do this...
Observe this...
Pass/Fail
1.
On Propaq M, install the airway adapter
set and sampling line in the connector at
the side of the unit.
2.
Press the
CO
2
quick access key
(
)
to
activate the CO
2
trace.
The CO
2
trace is on the display.
o
o
3.
On Propaq M, gently blow into the airway
adapter set while observing the display.
The CO
2
waveform changes on the
display.
o
o
CO
2
CO
2