![ZOLL Propaq M Service Manual Download Page 41](http://html1.mh-extra.com/html/zoll/propaq-m/propaq-m_service-manual_3993216041.webp)
Propaq M Error Messages
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
37
NIBP AIR LEAK
Major air leak or could not
reach target pressure.
Attempt to identify
problems with the hose
and/or cuff by doing the
following:
•
Check the hose
connection to the
device. Make sure that
the hose is tightly
screwed onto the NIBP
connector.
•
Ensure that the cuff is
appropriately wrapped
and is not too loose.
•
Inspect the cuff and
hose for leaks. Inspect
the O-ring on the NIBP
hose connector.
Consider replacing the
cuff and/or hose.
If the problem persists,
perform one of the following
procedures in
Chapter 1,
"NIBP Valve, Pump, and
Tubing Integrity Tests"
:
•
NIBP Leak Test
•
NIBP Inflation Test
Inspect internal NIBP
tubing for leaks.
NIBP EQUIPMENT CHECK
Device is unable to obtain a
reading.
Check the hose and cuff
connections. If problem
persists, perform one or
more of the following
procedures outlined in
Chapter 1, "NIBP Valve,
Pump, and Tubing Integrity
Tests"
:
•
NIBP Leak Test
•
NIBP Deflation Test
•
NIBP Inflation Test
NIBP WEAK PULSE
Too few pulses are detected
during the reading attempt.
Tighten the cuff and ensure
appropriate placement.
NIBP TEMP OUT OF
RANGE
Module temperature is out of
its specified operating range.
Move the device to a cooler
or warmer environment.
NIBP ARTIFACT
Unable to determine blood
pressure. Check
connections; restrict patient
movement.
Attempt to identify and
eliminate the source of
artifact. If problem persists,
replace the NIBP module.
Error Message
Explanation
Technical Action
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...