![ZOLL Propaq M Service Manual Download Page 57](http://html1.mh-extra.com/html/zoll/propaq-m/propaq-m_service-manual_3993216057.webp)
Removing the EtCO2 and NIBP Modules
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
53
9
Remove five (5) screws and washers from
the NIBP unit to release it.
Note: During reassembly, torque screws to
6 in-lbs.
10
Cut the tie wrap to the ferrite.
11
Remove multi-wire EtCO
2
cable with
ferrite.
12
Disconnect the brown SpO
2
flex cable as
shown.
13
Disconnect the NIBP tubing from the side
panel in the area illustrated.
tie wrap
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...