72
9650-0804-01 Rev. A
Bio-Tek Instruments BP Pump
Adult/Pediatric Mode
Neonate Mode
Bio-Tek Instruments BP Pump2
Adult/Pediatric Mode
Neonate Mode
FastBP
StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150
198
151
200
149
120/80
120
83
116
84
80/50
83
54
77
53
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
60/30
56
26
35/15
31
12
FastBP
StepBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
Expected
Systolic
Expected
Diastolic
200/150
202
157
201
146
120/80
118
80
116
82
80/50
86
51
77
55
NeoBP
Simulator
Setting
Expected
Systolic
Expected
Diastolic
80/50
78
46
60/30
57
31
35/15
31
12
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...