NIBP Valve, Pump, and Tubing Integrity Tests
9650-0804-01 Rev. A
ZOLL Propaq M Service Manual
11
NIBP Valve, Pump, and Tubing Integrity Tests
NIBP Leak Test
Equipment
Single lumen BP hose, a 500 cc volume, and a calibrated manometer.
Note:
If a 500 cc volume is not available, an Adult BP cuff wrapped around a rigid cylinder
(such as a glass jar) will suffice
Procedure
1. Power on the Propaq M.
2.
Connect the Propaq M to the single lumen hose and the 100 cc volume. Alternatively, replace
the 100 cc volume with a standard Adult cuff wrapped around a rigid cylinder, such as a glass
jar. If available, connect a calibrated manometer to the air line using a T-fitting.
3.
On the Propaq M, do the following to enter the NIBP Test control panel:
•
Press the Setup quick access key
(
).
•
Use the navigation keys to select
Supervisor
>
Service
>
Device Tests
>
NIBP Test
.
4.
From the Propaq M NIBP Test control panel, use the navigation keys to select:
•
Close Valve
•
Turn Pump On
5.
Monitor the pressure reported by NIBP Test control panel or the external pressure standard.
6.
Select
Turn Pump Off
as soon as the pressure reaches the target pressure of 250 +/- 5 mmHg.
7.
Allow the system to stabilize for 10 seconds.
8.
Select
Mark P1
to mark the starting pressure. If available, measure the starting pressure
reported by the external pressure standard. Record the starting pressure as P
S
.
9.
Wait 15 seconds.
10.
.Select
Mark P2
to mark the ending pressure. If available, measure the ending pressure
reported by the external pressure standard. Record the ending pressure as P
E
.
11.
From the NIBP Test control panel, select
Open Valve
to release the pressure.
12.
Calculate the leak rate: L = (P
s
- P
e
) / (15 seconds).
Note:
If using the pressure reported in the NIBP Test control panel, calculate the Leak Rate
using the Primary transducer pressures.
Acceptable Results
The unit passes the test if the system leaks less than 5 mmHg over the 15-second time interval
(Leak Rate is less than 1/3).
Summary of Contents for Propaq M
Page 1: ...Propaq M Service Manual 9650 0804 01 Rev A December 2010...
Page 36: ...CHAPTER 1 MAINTENANCE TESTS 32 9650 0804 01 Rev A...
Page 44: ...CHAPTER 2 TROUBLESHOOTING 40 9650 0804 01 Rev A...
Page 50: ...46 9650 0804 01 Rev A 11 Disconnect the NIBP tube from the rear case NIBP hoses NIBP tube...
Page 52: ...48 9650 0804 01 Rev A 6 Remove the ZIF Zero Insertion Force flex cable using an orange stick...
Page 66: ...CHAPTER 5 REPLACEMENT PARTS 62 9650 0804 01 Rev A...
Page 70: ...CHAPTER 6 FUNCTIONAL DESCRIPTION 66 9650 0804 01 Rev A...
Page 72: ...68 9650 0804 01 Rev A Propaq M Electrical Hardware Interconnect Diagram...
Page 74: ...70 9650 0804 01 Rev A...
Page 78: ...74 9650 0804 01 Rev A...